Remote control commands
R&S
®
RTO6
1384
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
TRIGger<m>:RFReject
<Bandwidth>
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Bandwidth>
Range:
100E+3 to 2E+9
Increment: 1000
*RST:
1E+6
Default unit: Hz
Usage:
Asynchronous command
24.8.11
Skew
....................................................................................... 1384
........................................................................................... 1384
.................................................................................... 1385
CHANnel<m>:SKEW:MANual
<ManualCompens>
If enabled, the skew offset value (
) is used for compensa-
tion. This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompens>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
CHANnel<m>:SKEW:TIME
<Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensa-
ted by the instrument automatically. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Acquisition and setup