Remote control commands
R&S
®
RTO6
1494
User Manual 1801.6687.02 ─ 05
FIRSt
Time is measured either between rising edges or between falling
edges. The first edge is taken for the measurement. In single
measurements, it works the same way as "Either".
Only available for analog measurement sources.
*RST:
FIRSt (analog source), POSitive (digital source)
Usage:
Asynchronous command
MEASurement<m>:AMPTime:PSLope
<PulsesSlope>
Sets the first slope of the pulses to be counted. The setting is only relevant for pulse
count measurement (
MEASurement<m>:MAIN PULCnt
or
MEASurement<m>:ADDitional PULCnt,ON
).
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<PulsesSlope>
POSitive | NEGative | EITHer
Count either positive or negative pulses, or both.
*RST:
POSitive
Usage:
Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:DIRection
<EdgeCntDirct>
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..2
Selects the source number.
Parameters:
<EdgeCntDirct>
FRFI | FRLA
FRFI - FRom FIrst, counting starts with the first edge of the
waveform.
FRLA - FRom LAst, counting starts with the last edge of the
waveform.
*RST:
FRFI
Usage:
Asynchronous command
Automatic measurements