Protocol analysis
R&S
®
RTO6
647
User Manual 1801.6687.02 ─ 05
●
Less than, Greater than
●
Less or equal, Greater or equal
●
In range, Out of range
Remote command:
Identifier min ← Serial Message setup
Defines the bit pattern of the message identifier.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
Remote command:
Identifier max ← Serial Message setup
The second identifier pattern is required to specify a range with conditions "In range"
and "Out of range".
Remote command:
Data Condition ← Serial Message setup
Selects the operator to set a specific data pattern or a data range.
The available operators:
●
Equal, Not equal
●
Less than, Greater than
●
Less or equal, Greater or equal
●
In range, Out of range
Remote command:
Data min ← Serial Message setup
Sets the data pattern. Enter the bytes in msb first bit order. The maximum pattern
length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
Remote command:
Data max ← Serial Message setup
The second data pattern is required to specify a range with conditions "In range" and
"Out of range".
Remote command:
SENT (option R&S
RTO6-K520)