![R&S RTO6 Скачать руководство пользователя страница 1452](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461452.webp)
Remote control commands
R&S
®
RTO6
1452
User Manual 1801.6687.02 ─ 05
REFCurve<m>:RESCale:HORizontal:STATe
<State>
Enables and disables the horizontal stretching.
Stretching changes the display of the waveform independent of the horizontal settings
of the source waveform and of the horizontal diagram settings.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
REFCurve<m>:RESCale:HORizontal:FACTor
<ScaleFactor>
A factor greater than 1 stretches the waveform horizontally, a factor lower than 1 com-
presses the curve.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<ScaleFactor>
Range:
1E-6 to 1E+6
Increment: 0.1
*RST:
1
Usage:
Asynchronous command
REFCurve<m>:RESCale:HORizontal:OFFSet
<Offset>
Moves the waveform horizontally. Positive values shift the waveform to the right, nega-
tive values shift it to the left.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<Offset>
Range:
-100E+24 to 100E+24
Increment: 0.01
*RST:
0
Default unit: s
Usage:
Asynchronous command
Waveform analysis