Remote control commands
R&S
®
RTO6
1515
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
12 | 13
12: Peak list measurements
13: Harmonic search measurements
Parameters:
<MaxNoOfResults>
Range:
1 to 1000
Increment: 1
*RST:
10
Usage:
Asynchronous command
MEASurement<m>:SPECtrum:RESult<n>:MODE
<ResultMode>
Sets the way how the measurement results are displayed. Available only if
MEASurement:MAIN
is set to
PLISt
or
HAR
.
Suffix:
<m>
.
1..10
<n>
12 | 13
12: Peak list measurements
13: Harmonic search measurements
Parameters:
<ResultMode>
ABS | REL
ABS
The harmonics/peaks are shown in absolute value, dBm.
REL
The level of the carrier is shown in absolute value dBm. The val-
ues the harmonics/peaks are shown relatively to the carrier in
dBc.
*RST:
ABS
Example:
:MEASurement1:SPECtrum:RESult13:MODE ABS
The harmonic measurements are shown in absolute value, dBm.
Usage:
Asynchronous command
MEASurement<m>:RESult:MAXCount
<Number>
MEASurement<m>:RESult:MAXCount?
<MeasType>
Defines the maximum number of peaks that are labeled in the diagram.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Automatic measurements