Acquisition and waveform setup
R&S
®
RTO6
174
User Manual 1801.6687.02 ─ 05
Bandwidth
Sets the bandwidth limit of all R&S
RT-ZVC probes. The bandwidth specifies the maxi-
mum frequency at which a purely sinusoidal signal is still transferred at 89 % (1 dB) of
its amplitude.
The bandwidth of some current channels is restricted to 300 KHz due to their vertical
settings.
Remote command:
Couple resolution
Sets the resolution of all R&S
RT-ZVC channels.
"On"
The resolution of the analog channels is applied to R&S
RT-ZVC
channels. The signal is automatically interpolated or decimated to get
the analog resolution.
"Off"
The resolution of R&S
RT-ZVC channels is set in a way so that the
record length of the waveforms is minimum 1000 samples.
Remote command:
5.7.1.2
Power calculation
Access: [App Cockpit]> "Analysis" > "Multi-channel probe" > "Power Cal." tab
State
Enables the corresponding channel of the probe. The number of available channels
depend on the characteristics of your multi-channel power probe.
Remote command:
R&S
RT-ZVC probe