Record length limit
Sets a limit for the record length to prevent very large records. This value is only avail-
able if "Auto adjustment" is on and a constant resolution is selected. If you increase the
time scale, the resolution remains constant and the record length increases until the
limit is reached. Further increase of the time scale changes the resolution and keeps
the record length limit.
See also:
●
●
Remote command:
5.4.2
Mode settings
Access: "Menu" > "Acquire" > "Mode" tab.
Decimation
Selects the decimation mode. Decimation reduces the data stream of the ADC to a
stream of waveform points with lower sample rate and a less precise time resolution.
The R&S
RTO6 uses decimation, if the waveform "Sample rate" is less than the ADC
sample rate. In this case, interpolation is not possible.
There are different methods to define the recorded waveform point out of n sample
points:
"Sample"
One of n samples in a sample interval of the ADC is recorded as
waveform point, the other samples are discarded. The time between
the two adjacent waveform points is exactly the resolution. Very short
glitches might remain undiscovered by this method.