CURSor<m>:MAXimum:NEXT
Sets cursor 2 to the next smaller peak (from the current position).
Suffix:
<m>
.
1..4
Selects the cursor set.
Usage:
Event
24.12
Automatic measurements
This chapter contains all remote commands to set up automatic measurements and to
analyze the measurement results.
Selection of the measurement group: MEASurement<m>
With R&S
RTO6, you can configure up to eight simultaneous measurement groups. In
manual operation, these eight measurement groups are represented by subtabs "MG1"
to "MG8" in the "Measurements" dialog box. For remote operation, the measurement
group is indicated by the suffix MEAS<m>, containing the number of the measurement
group.
Remote control: measurement suffix <m>
Manual operation: "MG" subtab
1 to 8 correspond to
"MG1" to "MG8"
9, 10: do not use
Not available
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands
*OPC
,
*OPC?
or
*WAI
can be used after the command or a command set.
For more information, see:
●
www.rohde-schwarz.com/rc-via-scpi
, chapter "Command Sequence and Synchro-
nization"
●
Chapter 24.3, "Programming examples"
............................................................................1489
Arithmetic for amplitude/time and spectrum measurements
...............................1499
............................................................................... 1508
........................................................................................ 1518
Automatic measurements