Remote control commands
R&S
®
RTO6
2396
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:DDR:DOPerator
<SearchName>,<Frame>, <Field>, <Operator>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:DOPerator
<SearchName>,<DataOperator>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:DOPerator?
<SearchName>
Sets the operator to set a specific data pattern to be searched in the selected field of
the selected frame.
Suffix:
<m>
.
*
<n>
*
Parameters:
<DataOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one endpoint
value to be set using
INRange | OORange
In range, out of range. These conditions require a range of end-
point values to be set using
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:DDR:FRENable
<SearchName>,<Frame>, <Enabler>
SEARch:TRIGger:DDR:FRAMe<m>:ENABle
<SearchName>,<Enable>
SEARch:TRIGger:DDR:FRAMe<m>:ENABle?
<SearchName>
Enables or disables the specific frame to be searched for.
Suffix:
<m>
.
*
Parameters:
<Enable>
ON | OFF
*RST:
OFF
Parameters for setting and query:
<SearchName>
The following frames are available:
Read
Write
Read NC
Write NC
Protocols