Measurements
R&S
®
RTO6
365
User Manual 1801.6687.02 ─ 05
Type
Selects and enables the limit or margin check.
"Off"
No limit check is performed.
"Limit only"
Limits are checked for violation.
"Margin &
Limit"
Margins and limits are checked for violation.
Upper limit, Lower limit, Upper margin, Lower margin, Valid range
Set the limits and margins for each measurement, and also specify the valid range.
Upper limit
Upper margin
Lower margin
Lower limit
within
outside
outside
upper
lower
Upper margin
Upper limit
Lower limit
Lower margin
Figure 8-15: Limit and margin definition
Limits are stricter than the margins for the value check. Thus, the margins must be
within the valid range. If necessary, the limit and margin values are adapted according
to the selected valid range.