Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1177
User Manual 1801.6687.02 ─ 05
19.5.3.2
Configuring switching loss
For details of the configuration settings, see
Chapter 19.5.3.3, "Switching loss set-
1. Select "Analysis" > "Power".
2. Under "Power Path", select "Switching Loss".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
4. Deskew the probes as described in
Chapter 19.1.1.1, "Auto deskew"
5. Connect the probes to the DUT as shown in the "Channels" tab:
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8. Enable the parameters you want to measure.
9. Select an"Optimize Scaling" option.
Power path