![R&S RTO6 Скачать руководство пользователя страница 1633](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461633.webp)
Remote control commands
R&S
®
RTO6
1633
User Manual 1801.6687.02 ─ 05
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:GATE:RELative:STARt
<SearchName>,<RelativeStart>
SEARch:GATE:RELative:STARt?
<SearchName>
Defines the starting value for the gate.
Parameters:
<RelativeStart>
Range:
0 to 100
Increment: 0.1
*RST:
0
Default unit: %
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:GATE:RELative:STOP
<SearchName>,<RelativeStop>
SEARch:GATE:RELative:STOP?
<SearchName>
Defines the end value for the gate.
Parameters:
<RelativeStop>
Range:
0 to 100
Increment: 0.1
*RST:
100
Default unit: %
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:GATE:ZCOupling
<SearchName>,<ZoomCoupling>
SEARch:GATE:ZCOupling?
<SearchName>
If enabled, the gate area is set to the limits of a zoom area.
The zoom diagramm is selected using
Parameters:
<ZoomCoupling>
ON | OFF
*RST:
OFF
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
Search