![R&S RTO6 Скачать руководство пользователя страница 1442](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461442.webp)
Remote control commands
R&S
®
RTO6
1442
User Manual 1801.6687.02 ─ 05
<ZoomName>
Name of the zoom diagram
Example:
See
Chapter 24.3.1.2, "Creating zoom diagrams"
Usage:
Asynchronous command
LAYout:ZOOM:HORZ:RELative:STARt
<DiagramName>,
<ZoomName>,<RelativeStart>
LAYout:ZOOM:HORZ:RELative:STARt?
<DiagramName>, <ZoomName>
Defines the lower limit of the zoom area on the x-axis.
Parameters:
<RelativeStart>
Range:
0 to 100
Increment: 0.1
*RST:
0
Default unit: %
Parameters for setting and query:
<DiagramName>
Name of the diagram on which the zoom area is based.
<ZoomName>
Name of the zoom diagram
Usage:
Asynchronous command
LAYout:ZOOM:HORZ:RELative:STOP
<DiagramName>,
<ZoomName>,<RelativeStop>
LAYout:ZOOM:HORZ:RELative:STOP?
<DiagramName>, <ZoomName>
Defines the upper limit of the zoom area on the x-axis.
Parameters:
<RelativeStop>
Range:
0 to 100
Increment: 0.1
*RST:
100
Default unit: %
Parameters for setting and query:
<DiagramName>
Name of the diagram on which the zoom area is based.
<ZoomName>
Name of the zoom diagram
Usage:
Asynchronous command
LAYout:ZOOM:VERTical:MODE
<DiagramName>, <ZoomName>,<Mode>
LAYout:ZOOM:VERTical:MODE?
<DiagramName>, <ZoomName>
Defines whether absolute or relative values are used to specify the y-axis values.
Since the zoom area refers to the active signal, relative values ensure that the zoom
area remains the same.
Waveform analysis