Remote control commands
R&S
®
RTO6
1908
User Manual 1801.6687.02 ─ 05
To trigger on an error event, select the corresponding trigger type with
.
Parameters:
<FormError>
ON | OFF
*RST:
OFF
TRIGger<m>:SENT:PULSeerror
<CalibPulseError>
Trigger son calibration pulse erors in transmission sequences.
An error occurs when
●
the duration of the "Calibration/Sync" pulse (in ticks) is less than 56*(1-clock toler-
ance) or more than 56*(1+clock tolerance)
●
the "Calibration/Sync" pulse duration of frame (n-1) varies by more than 1.5625%
from the "Calibration/Sync" pulse duration of frame (n)
To trigger on an error event, select the correspondig trigger type with
.
Parameters:
<CalibPulseError>
ON | OFF
*RST:
OFF
TRIGger<m>:SENT:PPERioderror
<PulsePeriodErr>
Triggers on pulse period errors.
An error occurs when a nibble has any of the following:
●
number of ticks at low is less than 4 ticks.
●
nibble value < 0 (less than 12 ticks) or > 15 (more than 27 ticks).
To trigger on an error event, select the correspondig trigger type with
.
Parameters:
<PulsePeriodErr>
ON | OFF
*RST:
OFF
TRIGger<m>:SENT:CRCerror
<CRCError>
Triggers on CRC errors in both, the transmission sequences and serial messages.
A CRC error occurs when the CRC calculated by the receiver differs from the received
value in the CRC sequence. The CRC length is 4 bits for transmission sequences and
short serial messages, and 6 bit of enhanced serial messages.
To trigger on an error event, select the correspondig trigger type with
.
Protocols