Remote control commands
R&S
®
RTO6
2403
User Manual 1801.6687.02 ─ 05
EYE<m>:TIMReference:PROTocol:SELect
..................................................................... 2407
................................................................................ 2407
EYE<m>:MEASurement:MAXPrecision
.......................................................................... 2408
EYE<m>[:STATe]
<State>
Activates or deactivates the eye diagram.
Suffix:
<m>
.
1..4
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
EYE<m>:SOURce
<MainSource>
Selects the waveform from which the eye diagram is generated (data source). For ana-
log channels, peak detect, envelope, and average waveforms are not supported.
Suffix:
<m>
.
1..4
Parameters:
<MainSource>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
Usage:
Asynchronous command
EYE<m>:HORizontal[:UNIT]
<TimeUnits>
Sets the method to define the horizontal scale of the eye diagram.
Suffix:
<m>
.
1..4
Parameters:
<TimeUnits>
ATIMe | UINTerval
ATIMe
Absolute time
UINTerval
Unit intervals. You can use unit interval, if the nominal frequency
of the clock signal is known.
*RST:
UINTerval
Usage:
Asynchronous command
EYE<m>:HORizontal:ABSolute:RANGe
<HorizontalRange>
Sets the time range that is covered by the eye diagram.
Protocols