Remote control commands
R&S
®
RTO6
1398
User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
100E-12
Default unit: s
Usage:
Asynchronous command
24.9.6
Window trigger
TRIGger<m>:LEVel<n>:WINDow:LOWer
........................................................................1398
TRIGger<m>:LEVel<n>:WINDow:UPPer
........................................................................ 1398
......................................................................................1398
......................................................................................... 1399
.......................................................................................1400
....................................................................................... 1400
TRIGger<m>:LEVel<n>:WINDow:LOWer
<Level>
TRIGger<m>:LEVel<n>:WINDow:UPPer
<Level>
Set the lower and upper voltage limits for the window.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n>
1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Parameters:
<Level>
Range:
-10 to 10
Increment: 1E-3
*RST:
Lower = -0.1, upper =0.1
Default unit: V
Usage:
Asynchronous command
TRIGger<m>:WINDow:RANGe
<RangeMode>
Defines the signal run in relation to the window:
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode>
ENTer | EXIT | WITHin | OUTSide
Trigger