Remote control commands
R&S
®
RTO6
1889
User Manual 1801.6687.02 ─ 05
Parameters:
<DataPositionTo>
Range:
0 to 255
Increment: 1
*RST:
0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:FCONdition
<SearchName>,<IdOperator>
SEARch:TRIGger:FLXRay:FCONdition?
<SearchName>
Sets the operator to set a frame ID or a frame ID range.
Parameters:
<IdOperator>
OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The frame ID is not relevant for the search condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one frame ID to
be set with
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with
and
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:FMAX
<SearchName>,<IdPatternTo>
SEARch:TRIGger:FLXRay:FMAX?
<SearchName>
Sets the the end value of an identifier range if the condition
is set to
INRange
or
OORange
.
Parameters:
<IdPatternTo>
Numeric or string pattern, see
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:FMIN
<SearchName>,<IdPattern>
SEARch:TRIGger:FLXRay:FMIN?
<SearchName>
Specifies a frame identifier pattern - the number of the slot - or sets the the start value
of an identifier range.
Protocols