![R&S RTO6 Скачать руководство пользователя страница 1659](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461659.webp)
Remote control commands
R&S
®
RTO6
1659
User Manual 1801.6687.02 ─ 05
24.16.7
Long term measurement results and measurement histogram
export to file
.......................................................................................1659
.........................................................................................1659
........................................................................................1659
.........................................................................................1660
....................................................................................... 1660
EXPort:MEASurement:SELect
<SelcMeas>
Selects the measurement for export of long term or measurement histogram data.
Parameters:
<SelcMeas>
MEAS1 | MEAS2 | MEAS3 | MEAS4 | MEAS5 | MEAS6 |
MEAS7 | MEAS8
*RST:
MEAS1
Example:
See
"Exporting long-term measurement data to file"
Usage:
Asynchronous command
EXPort:MEASurement:TYPE
<ExportType>
You can export the result data of the long term measurement, or the measurement his-
togram, or the track data.
To export the measurement histogram, it must be enabled using
.
To export the long term results, the long term measurement must be enabled using
.
To export a track, the track must be enabled before. Track measurements require an
option, see
Parameters:
<ExportType>
LONGterm | HISTogram | TRACk
LONGTERM = LONGterm, HISTOGRAM = HISTogram
*RST:
HISTogram
Example:
See
"Exporting long-term measurement data to file"
Usage:
Asynchronous command
EXPort:MEASurement:NAME
<Path>
Sets the file name and path to save the long term or measurement histogram data to.
Data management