Protocol analysis
R&S
®
RTO6
883
User Manual 1801.6687.02 ─ 05
Virtual Channel
Triggers on a specific virtual channel, an independent data stream for up to four periph-
erals.
Remote command:
Data Type
Sets the specified data type to be triggered on. The data type setup consists of the
condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:DPHY:DTYPe:CONDition
Word Count
Sets the specified specific word count to be triggered on. The word count setup con-
sists of the condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:DPHY:WCOunt:CONDition
Data Value
Set the specified data value to be triggered on. The data value setup consists of the
condition and one or two data value patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
D-PHY (option R&S
RTO6-K580)