![R&S RTO6 Скачать руководство пользователя страница 339](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646339.webp)
Measurements
R&S
®
RTO6
339
User Manual 1801.6687.02 ─ 05
If "Peak excursion" is 30
dB, the peaks 1 to 5 are found. If "Peak excursion" is 20
dB,
also the peaks 6 to 10 are found. The cursor position is on peak 6.
Remote command:
MEASurement<m>:SPECtrum:PEXCursion
Max results
Sets the maximum number of measurement results that are listed in the result table.
Available for peak list and harmonic search measurements.
Remote command:
MEASurement<m>:SPECtrum:RESult<n>:COUNt
Result mode
Selects the the way the measurement results are displayed. Available for peak list and
harmonic search measurements.
"Absolute"
The harmonics/peaks are shown in absolute value, dBm.
"Relative"
The level of the carrier is shown in absolute value dBm. The values
the harmonics/peaks are shown relatively to the carrier in dBc.
Remote command:
MEASurement<m>:SPECtrum:RESult<n>:MODE
Result labels
For peak lists only, requires option R&S
RTO6-K37: Use labels to describe the detec-
ted peaks in the spectrum diagram.
You can configure what is displayed in the labels. To change the label design, use
"Menu" > "Settings" > "Appearance" > "Peak List", see
"Enable"
Displays a description for each detected peak in the spectrum dia-
gram.
"Max. peak
count"
Defines the maximum number of peaks that are labeled in the dia-
gram. The result table lists all peaks.
"Show Fre-
quency"
Includes the frequency of the detected peak in the diagram labels.
Remote command:
MEASurement<m>:RESult:SHLabels
MEASurement<m>:RESult:LABorder
MEASurement<m>:RESult:MAXCount
MEASurement<m>:RESult:SHFRequency
Automatic measurements