Remote control commands
R&S
®
RTO6
2438
User Manual 1801.6687.02 ─ 05
Parameters:
<Bit>
HIGH | LOW | DONTCARE | DONTcare
Bit value: 1 (HIGH), 0 (LOW), or X (DONTCARE = DONTcare)
Usage:
Asynchronous command
TRIGger<m>:PARallel:PATTern:MODE
<Mode>
Sets the mode of the timing condition.
Parameters:
<Mode>
OFF | TIMeout | WIDTh
OFF
No timing condition, only the logical pattern condition is relevant.
TIMeout
Defines a minimum time qualification to avoid triggering on
unstable or transitional conditions. Use
PARallel:PATTern:TIMeout[:TIME]
WIDTh
Sets a pulse width as timing condition. The pulse starts when
the pattern comes true, and the trigger event occurs when the
pattern comes false during the specified time limit. Use
TRIGger<m>:PARallel:PATTern:WIDTh:RANGe
TRIGger<m>:PARallel:PATTern:WIDTh[:WIDTh]
TRIGger<m>:PARallel:PATTern:WIDTh:DELTa
the width.
*RST:
OFF
Usage:
Asynchronous command
TRIGger<m>:PARallel:PATTern:TIMeout:MODE
<TimeoutMode>
Sets the state condition for the timeout qualification if
is set to TIMeout. To set the time limit, use
.
Parameters:
<TimeoutMode>
HIGH | LOW | EITHer
HIGH: The pattern stays true for the specified time.
LOW: The pattern stays false for the specified time.
EITHER: The pattern remains unchanged for the specified time.
*RST:
HIGH
Usage:
Asynchronous command
TRIGger<m>:PARallel:PATTern:TIMeout[:TIME]
<Time>
Defines the time limit for the timeout at which the instrument triggers.
Mixed signal option (MSO, R&S
RTO6-B1)