Remote control commands
R&S
®
RTO6
1946
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:CXPI:DATA:DCONdition
<SearchName>,<DataOperator>
SEARch:TRIGger:CXPI:DATA:DCONdition?
<SearchName>
Sets the condition for the data pattern. You can define no value, an exact value or a
value range.
Parameters:
<DataOperator>
OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
Any data pattern is considered
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
SEARch:TRIGger:CXPI:DATA:DCONdition
*RST:
OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:DATA:DMAX
<SearchName>,<DataMax>
SEARch:TRIGger:CXPI:DATA:DMAX?
<SearchName>
Sets the end value of a data type range if
is set to
INRange
or
OORange
.
Parameters:
<DataMax>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:DATA:DMIN
<SearchName>,<DataMin>
SEARch:TRIGger:CXPI:DATA:DMIN?
<SearchName>
Specifies a data bit pattern, or sets the start value of a pattern range.
Parameters:
<DataMin>
Parameters for setting and query:
<SearchName>
Protocols