![R&S RTO6 Скачать руководство пользователя страница 1616](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461616.webp)
Remote control commands
R&S
®
RTO6
1616
User Manual 1801.6687.02 ─ 05
<SignalSource>
C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 | C3W2 |
C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | TRK1 | TRK2 |
TRK3 | TRK4 | TRK5 | TRK6 | TRK7 | TRK8 | Z1V1 | Z1V2 |
Z1V3 | Z1V4 | Z1I1 | Z1I2 | Z1I3 | Z1I4 | Z2V1 | Z2V2 | Z2V3 |
Z2V4 | Z2I1 | Z2I2 | Z2I3 | Z2I4
Source of the search, see
Chapter 24.4.2, "Waveform parame-
Usage:
Asynchronous command
24.15.8
Timeout search conditions
Trigger level setting:
................................................................................1616
................................................................................... 1616
SEARch:TRIGger:TIMeout:RANGe
<SearchName>,<TimeoutMode>
SEARch:TRIGger:TIMeout:RANGe?
<SearchName>
Selects the relation of the signal level to the trigger level:
Parameters:
<TimeoutMode>
HIGH | LOW | EITHer
HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST:
HIGH
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:TIMeout:TIME
<SearchName>,<Time>
SEARch:TRIGger:TIMeout:TIME?
<SearchName>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
100E-9
Default unit: s
Search