![R&S RTO6 Скачать руководство пользователя страница 2465](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462465.webp)
Remote control commands
R&S
®
RTO6
2465
User Manual 1801.6687.02 ─ 05
PSRC:REPRate
<RepetitionRate>
Sets the pulse frequency, the repetition rate of the pulse.
Parameters:
<RepetitionRate>
Available values depend on the selected clock mode:
Locked: 5/10/20/50/100/200/500 Hz, 1/5/10/25/50/100/250 MHz
Free running: 5/10/20/50/100/200/500 Hz, 1/5/10/25/50 MHz
*RST:
1E+6
Default unit: Hz
Usage:
Asynchronous command
PSRC:DUTYcycle
<DutyCycle>
Sets the duty cycle of the pulse, which is the ratio of the positive pulse width to the
period of the signal.
Parameters:
<DutyCycle>
For repetition rates >
5
MHz, the value is fixed at 50
%.
Range:
10 to 90
Increment: 10
*RST:
50
Default unit: %
Usage:
Asynchronous command
PSRC:CLOCkmode
<ClockMode>
Sets the dependency of the pulse clock on the instrument's reference clock.
Parameters:
<ClockMode>
LOCKed | FREerunning
LOCKed
The pulse source is locked to the reference clock of the instru-
ment.
FREerunning
The clock of the pulse source is independent. Deviations of the
system do not affect the pulse clock, and deterministic condi-
tions are avoided.
*RST:
FREerunning
Usage:
Asynchronous command
PSRC:RST
Resets the pulse source to the default state.
Usage:
Event
Asynchronous command
Pulse Source and TDR/TDT Analysis (options R&S