Remote control commands
R&S
®
RTO6
1757
User Manual 1801.6687.02 ─ 05
For triggering on the serial bus when the NRZ unclocked coding standard is selected,
analog channel sources are required.
Otherwise, if no serial bus trigger has been selected, permitted source selections
include the mathematical channels and the reference channels.
Suffix:
<m>
.
1..4
Parameters:
<SourceNRZData>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
*RST:
C1W1
BUS<m>:CMSB:NRZ:IDLPolarity
<PolarityIdle>
Selects the idle polarity of the custom serial bus data signal (only available for the cod-
ing standard "NRZ Unclocked").
Suffix:
<m>
.
1..4
Parameters:
<PolarityIdle>
IDLLow | IDLHigh
IDLLow
Sets the base value of the data bus to be "0". After an idle
period, the data signal starts with a low-to-high transition
IDLHigh
Sets the base value of the data bus to be "1". After an idle
period, the data signal starts with a high-to-low transition
*RST:
IDLLow
BUS<m>:CMSB:NRZ:CPOLarity
<PolarityClock>
Selects the polarity of the clock signal for the coding standard NRZ Clocked.
Suffix:
<m>
.
1..4
Parameters:
<PolarityClock>
IDLLow | IDLHigh
IDLLow
Sets the base value of the clock to be "0", the clock signal starts
with a low-to-high transition
IDLHigh
Sets the base value of the clock to be "1", the clock signal starts
with a high-to-low transition.
*RST:
IDLLow