![R&S RTO6 Скачать руководство пользователя страница 472](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646472.webp)
Data and file management
R&S
®
RTO6
472
User Manual 1801.6687.02 ─ 05
In this dialog, you can select an eye diagram that you want to export.
Eye export saves the accumulated values at each pixel of an eye diagram.
The export supports *.csv and *.bin files. Each export generates a header file and a
data file.
Content and format of the *.csv files
The header file contains the following values:
●
Name of the exported eye (e.g. Eye 1)
●
Corresponding source (e.g. C1W1)
●
Timestamp
●
Number of lines and columns
●
Vertical / horizontal range
The data file contains the comma separated list of the pixel values. The number of
lines and columns corresponds to the number of lines and rows of the eye diagram
image.
Content and format of the *.bin files
The header file contains the following values:
●
Name of the exported eye (e.g. Eye 1)
●
Corresponding source (e.g. C1W1)
●
Timestamp
●
Number of lines and columns
●
Vertical / horizontal range
The data file contains the following information in the given order
●
Number of bytes per value - uint32
●
Number of samples – uint32
●
Column wise sorted pixel values – ‘number of samples’ * ‘number of bytes per val-
ue’
Select/ Source
Selects the eye measuremen to be exported and displys its source.
Save and recall waveform data and results