![R&S RTO6 Скачать руководство пользователя страница 2041](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462041.webp)
Remote control commands
R&S
®
RTO6
2041
User Manual 1801.6687.02 ─ 05
BUS<m>:MDIO:DATA:SOURce
<SourceData>
Selects the source for the data signal. Permitted selections are the analog channels
"C1"
–
"C4" and the digital channels "D0"
–
"D15", but not the same as for "Clock".
Suffix:
<m>
.
1..4
Parameters:
<SourceData>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See
Chapter 24.4.2, "Waveform parameter"
*RST:
C1W1
BUS<m>:MDIO:CLOCk:THReshold:HIGH
<ThresClkHigh>
Defines the upper threshold level for the clock signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkHigh>
Range:
-5 to 5
Increment: 0.1
*RST:
2
Default unit: V
BUS<m>:MDIO:CLOCk:THReshold:LOW
<ThresClkLow>
Defines the lower threshold level for the clock signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkLow>
Range:
-5 to 5
Increment: 0.1
*RST:
0.8
Default unit: V
BUS<m>:MDIO:DATA:THReshold:HIGH
<ThresDatHigh>
Defines the upper threshold level for the data signal.
Suffix:
<m>
.
1..4
Protocols