![R&S RTO6 Скачать руководство пользователя страница 659](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646659.webp)
Protocol analysis
R&S
®
RTO6
659
User Manual 1801.6687.02 ─ 05
<sb:UNIT></sb:UNIT>
</sb:SIGNAL>
<sb:SIGNAL ID="Channel_2">
<sb:SHORT-NAME>Counter</sb:SHORT-NAME>
<sb:DESCRIPTION>""</sb:DESCRIPTION>
<sb:BIT-POSITION>11</sb:BIT-POSITION>
<sb:BIT-LENGTH>8</sb:BIT-LENGTH>
<sb:BYTE-ORDER>MSB</sb:BYTE-ORDER>
<sb:FACTOR>1.0</sb:FACTOR>
<sb:OFFSET>0.0</sb:OFFSET>
<sb:MIN>0</sb:MIN>
<sb:MAX>10000.0</sb:MAX>
<sb:UNIT></sb:UNIT>
</sb:SIGNAL>
</sb:SIGNALS>
</sb:FRAME>
</sb:FRAMES>
</sb:LABEL-LIST-FILE>
13.10.5
SENT decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1.
In the "Protocol" dialog > "Setup" tab, enable "Decode".
2.
In the "Protocol" dialog > "Display" tab, select additional result display settings:
"Show decode table" and "Show binary signals".
For a description of the display settings, see also
The instrument captures and decodes the signal according to the standard definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
Data is decoded and displayed in the order of its reception. The "Decode results" box
shows the detailed decoded data for each frame as it is received.
SENT (option R&S
RTO6-K520)