![R&S RTO6 Скачать руководство пользователя страница 2200](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462200.webp)
Remote control commands
R&S
®
RTO6
2200
User Manual 1801.6687.02 ─ 05
Parameters:
<Enable>
ON | OFF
*RST:
ON
Example:
Trigger on CRC and length errors:
TRIGger:USBThree:ERRor1:ENABle ON
Enables CRC Error.
TRIGger:USBThree:ERENable "Length Error",ON
Enables Length Error.
TRIGger<m>:USBThree:FRENable
<Frame>, <Enabler>
TRIGger<m>:USBThree:FRAMe<n>:ENABle
<Enable>
Enables or disables the checking condition for the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n>
*
Specifies the frame number.
Parameters:
<Enable>
ON | OFF
*RST:
OFF
Example:
TRIGger:USBThree:FRAMe1:ENABle ON
Enables the checking condition for the TSEQ frame.
TRIGger:USBThree:FRENable "TS1" ON
Enables the checking condition for the TS1 frame.
TRIGger<m>:USBThree:FIENable
<Frame>, <Field>, <Enabler>
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:ENABle
<CondEnabler>
Enables or disables the checking condition for the selected field of the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n>
*
Specifies the frame number.
<o>
*
Specifies the field number within the frame.
Parameters:
<CondEnabler>
ON | OFF
ON
Checking condition enabled
OFF
Checking condition disabled
*RST:
OFF
Protocols