Remote control commands
R&S
®
RTO6
1529
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<HistgRef>
String with the name of the histogram
Example:
See
"Creating and reading histograms"
Usage:
Asynchronous command
MEASurement<m>:HISTogram:PROBability:TYPE
<HistgPrbDomMark>
Defines the marker reference in the probability domain.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<HistgPrbDomMark> PEAK | UPPK | LWPK | MAXimum | MINimum | MEDian | MEAN
PEAK
The y-value with the maximum sample value in the histogram
UPPK
The y-value at the maximum sample value in the upper half of
the histogram
LWPK
The y-value at the maximum sample value in the lower half of
the histogram
MAXimum
The highest y-value with a probability > 0
MINimum
The lowest y-value with a probability > 0
MEDian
The y-value for which half the samples lie above, the other half
below in the histogram
MEAN
The weighted arithmetic average of the histogram
*RST:
PEAK
Usage:
Asynchronous command
MEASurement<m>:HISTogram:PROBability:LIMit
<HistgPrbDomLim>
Defines a range around the probability marker.
Automatic measurements