Remote control commands
R&S
®
RTO6
1358
User Manual 1801.6687.02 ─ 05
Parameters:
<VoltageAdjust>
Correction voltage
Range:
-100E+24 to 100E+24
Increment: 1E-3
*RST:
0
Default unit: V
Usage:
Asynchronous command
24.8.7.6
Modular probes, deembedding
The commands of this chapter are relevant for R&S
RT
‑
ZM modular probes when used
for measurements on signal frequencies of 4
GHz and higher. The commands are
available in firmware version 4.00 and higher.
See also:
...............................................................................1358
PROBe<m>:DEEMbedding:TIPModule[:SELect]
............................................................. 1358
PROBe<m>:DEEMbedding:TIPModule:CSPFile
..............................................................1359
PROBe<m>:DEEMbedding:TIPModule:ZMA<n>:SUBModule
...........................................1359
PROBe<m>:DEEMbedding:BANDwidth
..........................................................................1359
................................................................................ 1360
PROBe<m>:DEEMbedding:REMProbeload
.................................................................... 1360
PROBe<m>:DEEMbedding:LOAD:IMPedance
................................................................ 1361
PROBe<m>:DEEMbedding:PMODe
<MeasMode>
Suffix:
<m>
.
1..4
Parameters:
<MeasMode>
DMODe | CMODe | PMODe | NMODe
See
*RST:
DMODe
Usage:
Asynchronous command
PROBe<m>:DEEMbedding:TIPModule[:SELect]
<ProbeTip>
Selects the tip module that is used for measurement.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<ProbeTip>
NONE | ZMA10 | ZMA12 | ZMA14 | ZMA15 | ZMA30 | ZMA40 |
ZMA50 | CUST
*RST:
ZMA10
Usage:
Asynchronous command
Acquisition and setup