![R&S RTO6 Скачать руководство пользователя страница 972](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646972.webp)
Protocol analysis
R&S
®
RTO6
972
User Manual 1801.6687.02 ─ 05
Decode Results Table
The results are shown in two tables:
●
"Decode results": contains information about all decoded frames
●
"Details frame": contains more detailed information about the selected frame in the
"Decode results" table.
Table 13-51: Content of the Decode result table
Column
Description
Frame
Frame count
State
Overall state of the frame indicating, for example, if the frame is valid or invalid
Start
Time of frame start in relation to the trigger point
Frame Type
Frame identifier specifying the data or control frame name
Word count
Total number of words in the frame
Table 13-52: Content of the Details frame table
Column
Description
State
State of the field
Numeric
Hexadecimal value of the field
Formatted
Formatted content of the field
Remote commands
Remote commands are described in
Chapter 24.17.28.2, "Decode results"
13.27.4
Eye diagram setup
This chapter describes the configuration of eye diagrams.
The configuration of an eye diagram mainly consists of the selection and setup of the
input source and timing reference source. You can also adjust display settings and
restrict the segments that contribute to the eye generation by various means.
You can configure and analyze up to four independent eye diagrams in parallel.
The "Eye Diagram" dialog box provides four tabs with all settings that are required to
configure eye diagrams.
DDR (option R&S
RTO6-K91 )