Protocol analysis
R&S
®
RTO6
555
User Manual 1801.6687.02 ─ 05
Autoset
Starts software algorithms for determining the signal threshold levels and bitrate.
Remote command:
Check bit rate
Enables the bit rate settings for the coding standards "Manchester" and "Manchester
II". This setting is not available for "NRZ Clocked", but always enabled for the coding
standard "NRZ Unclocked", and also for triggering on signals in any coding standard.
Remote command:
Bit Rate
Defines the transmission speed setting for the data signal. A bit rate definition is
optional for the coding standards "Manchester" and "Manchester II", not available for
"NRZ Clocked", but obligatory for "NRZ Unclocked" (and also for triggering on signals
in any coding standard). Default bit rate is 10
Mbps, permitted bit rates range from
300
bps to 2
Gbps.
For more details on the bit rate, see
Chapter 13.6.1.1, "Special features of Manchester
Remote command:
Check gap time
Enables the gap time settings (always enabled for the coding standard "NRZ
Unclocked", and also for triggering on signals in any coding standard).
Remote command:
Min gap time
Specifies a minimum gap time (idle time or timeout) between two frames. A gap time
definition is optional for the coding standards "Manchester", "Manchester II" and "NRZ
Clocked", but obligatory for "NRZ Unclocked" (and also for triggering on signals in any
coding standard). Default gap time is 10
µs, permitted gap times range from 1
ns to
1
s.
Custom: Manchester / NRZ (option R&S
RTO6-K510)