Remote control commands
R&S
®
RTO6
1531
User Manual 1801.6687.02 ─ 05
<n>
1..18
Number of the histogram measurement, see
Parameters:
<UpperLimit>
Range:
-100 to 100
Increment: 10E-6
*RST:
0
Usage:
Asynchronous command
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:MARGin
<LowerMargin>
MEASurement<m>:HISTogram:LCHeck<n>:UPPer:MARGin
<UpperMargin>
Define the lower and upper margins for the limit check, respectively. Margins are not as
strict as limits and must be within the valid value range. The valid range is defined
using the
MEASurement<m>:HISTogram:LCHeck<n>:VALid
command.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
<n>
1..18
Number of the histogram measurement, see
Parameters:
<UpperMargin>
Range:
-100 to 100
Increment: 10E-6
*RST:
0
Usage:
Asynchronous command
24.12.9
Display
............................................................................... 1531
MEASurement<m>:DISPlay:RESults
..............................................................................1532
............................................................................... 1532
MEASurement<m>:DISPlay:HISTogram
......................................................................... 1532
MEASurement<m>:DISPlay:LEVels
<DisplayLevels>
If enabled, the reference levels used for the measurement are displayed in the dia-
gram.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<DisplayLevels>
ON | OFF
*RST:
OFF
Automatic measurements