Remote control commands
R&S
®
RTO6
1532
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
MEASurement<m>:DISPlay:RESults
<DisplayResult>
If enabled, the intermediate result linesare displayed in the measurement diagram.
Result lines mark the samples in the waveform that are used to obtain the measure-
ment result. These are, for example, maximum and minimum values, mean, cycle start
and cycle end, and others.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<DisplayResult>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
MEASurement<m>:DISPlay:STYLe
<DisplayStyle>
Selects the style in which the measurement waveform is displayed.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<DisplayStyle>
LINE | MARKer
LINE
The individual data points are connected by a line.
MARKer
Only the individual data points are displayed as markers.
*RST:
LINE
Usage:
Asynchronous command
MEASurement<m>:DISPlay:HISTogram
<DispHistg>
Displays a histogram for the source of the selected measurement.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<DispHistg>
ON | OFF
*RST:
OFF