Remote control commands
R&S
®
RTO6
2334
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<ThresholdDminus> Range:
-10 to 10
Increment: 1E-3
*RST:
0
Default unit: V
BUS<m>:EBTB:DPLus:SOURce
<SourceDplus>
Selects the D+ source of the provided single-ended signal, if
BUS<m>:EBTB:TYPE
SINGle
is set.
Suffix:
<m>
.
1..4
Parameters:
<SourceDplus>
NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4
*RST:
C1W1
Usage:
Asynchronous command
BUS<m>:EBTB:DPLus:THReshold
<ThresholdDplus>
Sets the high threshold (D +) of the signal, if
BUS<m>:EBTB:TYPE SINGle
.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdDplus>
Range:
-10 to 10
Increment: 1E-3
*RST:
0
Default unit: V
BUS<m>:EBTB:FCSY
<CustSync>
Sets a pattern value, if
BUS<m>:EBTB:SYNC CUS
.
Suffix:
<m>
.
1..4
Parameters:
<CustSync>
BUS<m>:EBTB:SCOunt?
Returns the symbol count for the selected serial bus, i.e. the number of symbols in the
present acquisition.
Protocols