Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1193
User Manual 1801.6687.02 ─ 05
In the "Settings" tab, you configure the transient response measurement parameters
and display settings.
Horizontal SMPS Frequency
Sets the SMPS switching frequency. Set the value according to your DUT.
Remote command:
Derived scale
Shows the scale of the displayed results. This value is calculated as the inverse of the
"SMPS switching frequency"
Edge trigger
Sets the properties of the trigger.
Source ← Edge trigger
Sets the source channel of the trigger.
Remote command:
Slope ← Edge trigger
Sets the edge type for the trigger event.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
Output