Protocol analysis
R&S
®
RTO6
574
User Manual 1801.6687.02 ─ 05
Error types: Item name, Enable
The table lists the error types you can filter on. Select the required errors in the
"Enable" column. Available are "CRC" error and "Parity" error.
Remote command:
BUS<m>:CMSB:FILTer:ERRor<n>:ENABle
Field name / Condition / Enable
This table lists the field numbers and names in the selected frame together with a sum-
mary of the user settings of checking conditions for each field and a checkbox to
enable the checking.
Select a field in the table to specify the checking conditions for this field in the "Data",
"BitState" and/or "Index" dialog (whichever applies). The condition is only applied, and
the "Condition" is only shown in the table, if "Enable" is checked.
Remote command:
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:ENABle
Bit State
Defines the bit state to be checked for the selected field. Permitted bit states are "1",
"0" or "X" (do not care).
Remote command:
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:BIT
Data
Defines for the selected field, how a data check is executed.
"Condition"
Defining specific data or a data range requires to set the operator to
one of the following conditions: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies data or sets the start value of a data range.
"Max"
Sets the end value of a data range, if "Condition" is set to "In range"
or "Out of range".
Remote command:
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:DMAX
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:DMIN
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:DOPerator
Index
Defines for the selected field, how an index check is executed.
This function is avalible only for array fields.
"Condition"
Defining a specific index or an index range requires to set the opera-
tor to one of the following conditions: equal, in range.
"Min"
Specifies the index or sets the start value of an index range.
"Max"
Sets the end value of an index range, if "Condition" is set to "In
range".
Custom: Manchester / NRZ (option R&S