"Within"
Triggers on pulse intervals inside a given range. The range is defined
by "Interv. width" and "
±
Delta".
"Outside"
Triggers on intervals outside a given range. The range definition is
the same as for "Within" range.
"Shorter"
Triggers on intervals shorter than the given "Interv. width".
"Longer"
Triggers on intervals longer than the given "Interv. width".
Remote command:
Interv. width
Defines the time between two pulses.
Remote command:
±Delta
Defines a range around the "Interval width" value.
Remote command:
6.3.9
Slew rate
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Slew rate"
The slew rate trigger is also known as transition trigger. It triggers if the transition time
from the lower to higher voltage level (or vice versa) is shorter or longer as defined, or
outside or inside a specified time range.
The slew rate trigger finds slew rates faster than expected or permissible to avoid over-
shooting and other interfering effects. It also detects slow edges violating the timing in
pulse series.