Remote control commands
R&S
®
RTO6
1375
User Manual 1801.6687.02 ─ 05
Parameters:
<VerticalScale>
Range:
1E-15 to 10E+24
Increment: 10E-6
*RST:
3
Default unit: V/div
Usage:
Asynchronous command
ZVC:Z<m>:V<n>:SKEW
<DeskewOffset>
Sets the skew offset value for the voltage probe channel. This is a delay value, that is
known from the circuit specifics but cannot be compensated by the instrument auto-
matically.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<DeskewOffset>
Range:
-6E-6 to 6E-6
Increment: 200E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
ZVC:Z<m>:V<n>[:STATe]
<State>
Enables the corresponding voltage channel of the probe.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
ZVC:Z<m>:ID:NAME?
Queries the name of the probe.
Suffix:
<m>
.
1..2
Return values:
<Name>
Usage:
Query only
Asynchronous command
Acquisition and setup