Remote control commands
R&S
®
RTO6
2566
User Manual 1801.6687.02 ─ 05
PROBe<m>:DEEMbedding:TIPModule[:SELect]
<ProbeTip>
Selects the tip module that is used for measurement.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<ProbeTip>
NONE | ZMA10 | ZMA12 | ZMA14 | ZMA15 | ZMA30 | ZMA40 |
ZMA50 | CUST
*RST:
ZMA10
Usage:
Asynchronous command
PROBe<m>:DEEMbedding:TIPModule:ZMA<n>:SUBModule
<Value>
Defines additional settings for some tip modules:
For the browser module R&S
RT
‑
ZMA30, measure the space between the pins, and
select the appropriate value.
If R&S
RT
‑
ZMA40 is selected, choose the used submodule: semi-rigid cables or none.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
<n>
30 (R&S
RT
‑
ZMA30) or 40 (R&S
RT
‑
ZMA40)
Parameters:
<Value>
D16 | D25 | D45 | D75 | NONE | SRCS
For R&S
RT
‑
ZMA30:
D16 = 1.6 mm
D25 = 2.5 mm
D45 = 4.5 mm
D75 = 7.5 mm
For R&S
RT
‑
ZMA40:
NONE = no submodule
SRCS = semi-rigid cables
Usage:
Asynchronous command
PROBe<m>:DEEMbedding:BANDwidth
<ProbeBandwidth>
Sets the maximum bandwidth until which the signal is corrected. This maximum value
is the minimum bandwidth value of probe, probe tip and oscilloscope bandwidth, and
can not be higher than the highest frequency in a used S-parameter file. Consider that
most tip modules support the full bandwidth of the probe amplifier, but some tip mod-
ules have limited bandwidth.
Deembedding (option R&S
RTO6-K121)