![R&S RTO6 Скачать руководство пользователя страница 24](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646024.webp)
Getting started
R&S
®
RTO6
24
User Manual 1801.6687.02 ─ 05
–
Environments that are directly connected to a low-voltage supply network that
supplies residential buildings
●
Class
A equipment is intended for use in industrial environments. It can cause
radio disturbances in residential environments due to possible conducted and radi-
ated disturbances. It is therefore not suitable for class B environments.
If class
A equipment causes radio disturbances, take appropriate measures to
eliminate them.
3.1.4
Setting up the product
When setting up the instrument, follow the safety instructions:
●
●
3.1.4.1
Placing the product on a bench top
For standalone operation, place the instrument on a horizontal bench with even, flat
surface. The instrument can be used in horizontal position, standing on its feet, or with
the support feet on the bottom extended.
To place the product on a bench top
1. Place the product on a stable, flat and level surface.
2.
CAUTION!
The top surface of the product is too small for stacking. If you stack
another product on top of the product, the stack can fall over and cause injury.
If you want to save space, mount several products in a rack.
3.
CAUTION!
Foldable feet can collapse. See
Always fold the feet completely in or out. With folded-out feet, do not place any-
thing on top or underneath.
4.
NOTICE!
Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10
cm between the fan openings of the product
and any object in the vicinity.
Preparing for use