Protocol analysis
R&S
®
RTO6
600
User Manual 1801.6687.02 ─ 05
For a description of DBC files, see
Chapter 13.7.2.3, "Display settings"
13.7.6.1
Symbolic trigger
The "Symbolic" trigger type is available if a DBC label list file is loaded and applied. It
allows you to trigger on a specific data message, or a signal and its value that appears
inside the message.
Access:
1. Press [Bus].
2. In the "Setup" tab, set "Protocol = CAN/ CAN-FD".
3. In the "Shortcuts" tab, select "Setup trigger".
4. In the trigger dialog, set the "Type" = "Symbolic".
The "Frame type" is automatically set to "Data", and the "Identifier" is shown for infor-
mation.
Specific settings for the symbolic trigger are:
CAN and CAN FD (option R&S
RTO6-K520)