Protocol analysis
R&S
®
RTO6
910
User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:MPHY:SSIC:DOPerator
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:DOPerator
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:DMIN
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:DMAX
BitState
Defines the bit state to be checked for the selected field. Permitted bit states are "1",
"0" or "X" (do not care).
Remote command:
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:BIT
Index
Defines for the selected field, how an index check is executed.
"Condition"
Defining a specific index or an index range requires to set the opera-
tor to one of the following conditions: equal, in range.
"Min"
Specifies the index or sets the start value of an index range.
"Max"
Sets the end value of an index range, if "Condition" is set to "In
range".
Remote command:
TRIGger<m>:MPHY:SSIC:IOPerator
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:IOPerator
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:IMIN
TRIGger<m>:MPHY:SSIC:FRAMe<n>:FLD<o>:IMAX
13.24.3.4
Triggering on SSIC
Prerequisite: A bus is configured for the "M-PHY" > "Variant" = "SSIC" signal to be ana-
lyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus] key and selet the "Shortcuts" tab.
2. Press "Setup trigger".
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Tap "Serial bus" and select the serial bus that is set to SSIC.
5. If you trigger on errors, select the error types you want to find.
6. Select the frame type to be used for SSIC protocol analysis.
7. For some frame types, you can define the frame fields.
M-PHY and USB SSIC (option R&S
RTO6-K580 and K570)