Remote control commands
R&S
®
RTO6
2427
User Manual 1801.6687.02 ─ 05
<n>
0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Deskew>
Range:
-200E-9 to 200E-9
Increment: 200E-12
*RST:
0
Default unit: s
BUS<m>:PARallel:DESoffset
<DeskewOffset>
Sets a general delay for all digital channels. The resulting deskew of a digital channel
is the sum of this general value and the individual value set with
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
Parameters:
<DeskewOffset>
Range:
-200E-9 to 200E-9
Increment: 200E-12
*RST:
0
Default unit: s
BUS<m>:PARallel:BIT<n>:LABel
<Label>
Sets a name for the indicated digital channel. The name is displayed in the diagram.
For the parallel bus 1, you can also use
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
<n>
0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Label>
String containing the channel name
BUS<m>:PARallel:DATA:FORMat
<DataFormat>
Sets the number format for decoded data values on the display for the indicated paral-
lel bus.
It sets also the number representation for remote data transfer with
.
Suffix:
<m>
.
1..4
Mixed signal option (MSO, R&S
RTO6-B1)