"Time Code"
Sets the trigger to a time-code control code. You can define the data
pattern of the time code to be triggered on.
"Errors"
Triggers on an enabled error type.
Remote command:
Trigger Control Type
Triggers on a specific control type character.
"ANY"
Any control type character
"FCT"
Flow control token character
"EOP"
Normal end of packet character
"EEP"
Error end of packet character
Remote command:
Data (Time Code)
Sets the specified data type for the time code to be triggered on. The trigger data setup
consists of a comparison condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:SWIRe:TIME:CONDition
Data (Data Pattern)
Sets the specified data type for the data pattern to be triggered on. The data type setup
consists of the condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
SpaceWire (option R&S
RTO6-K530)