![R&S RTO6 Скачать руководство пользователя страница 942](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646942.webp)
Protocol analysis
R&S
®
RTO6
942
User Manual 1801.6687.02 ─ 05
"Damping"
Sets the damping factor, which is only relevant for second order PLL.
Remote command:
Initial phase sync.
Defines the phase reference for the first clock edge.
"First sample"
The first clock edge matches the first sample of the waveform at the
left border of the display.
"First data
edge"
The first clock edge matches the first edge of the data signal.
Remote command:
Selected results
The PLL requires some time to synchronize to the phase of the data stream. You can
select when the CDR algorithm returns clock edges:
"After initial
sync."
The clock edges of the synchronization time are discarded; results
are gathered after initial synchronization of the CDR. Thus, meaning-
ful TIE measurement results can be obtained.
"All"
All clock edges are used.
Remote command:
Software CDRs
Couples the protocol CDR to the selected software CDR. For more information about it
see,
Chapter 18.3, "Clock data recovery"
13.26.4
PCIe trigger
13.26.4.1
PCIe trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
PCIe (option R&S
RTO6-K72)