![R&S RTO6 Скачать руководство пользователя страница 724](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646724.webp)
Protocol analysis
R&S
®
RTO6
724
User Manual 1801.6687.02 ─ 05
"ANY"
Any control type character
"FCT"
Flow Control Token character
"EOP"
Normal End of Packet character
"EEP"
Error End of Packet character
Remote command:
Data (Time Code)
Sets the specified data type for the time code to be searched for. The data type setup
consists of the condition and one or two data patterns.
"Condition"
Defines the operator to set a specific data type ("Equal" or "Not
equal") or a data type range.
"Data Min/
Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:SWIRe:TIME:CONDition
Data (Data Pattern)
Sets the specified data type for the data pattern to be searched for. The data type
setup consists of the condition and one or two data patterns.
"Condition"
Defines the operator to set a specific data type ("Equal" or "Not
equal") or a data type range.
"Data Min/
Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:SWIRe:DATA:CONDition
Parity Error
Checks the parity of every frame and searches for even parity.
Remote command:
SEARch:TRIGger:SWIRe:ERRor:PARity
SpaceWire (option R&S