![R&S RTO6 Скачать руководство пользователя страница 769](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646769.webp)
Protocol analysis
R&S
®
RTO6
769
User Manual 1801.6687.02 ─ 05
Example: RFFE label list file
# Labels for RFFE protocol
# Column order: SID, SID text, Address, Address text
# ---------------------------------------------------------------------------
@PROTOCOL_NAME = rffe
0x01,LM8335
0x01,LM8335,0x00,CMTL_REG
0x01,LM8335,0x01,GPO_PULL_DIR
0x01,LM8335,0x02,GPO_PULL_ENABLE
0x01,LM8335,0x03,GPO_OUT_HIGH_CFG
0x01,LM8335,0x04,GPO_OUT_MASK
0x01,LM8335,0x05,GPO_OUT_DATA
0x01,LM8335,0x1C,PM_TRIG
0x01,LM8335,0x1D,PROD_ID
0x01,LM8335,0x1E,MAN_ID
0x01,LM8335,0x1F,USID_REG
0x05,LM3279
0x05,LM3279,0x00,VSET_CTRL
0x05,LM3279,0x01,STATE_CTRL
0x05,LM3279,0x02,GPO_CTRL
0x05,LM3279,0x1C,PM_TRIG
0x05,LM3279,0x1D,PROD_ID
0x05,LM3279,0x1E,MAN_ID
0x05,LM3279,0x1F,USID_REG
0x0A,Test
See also:
Chapter 13.1.4, "Content and format of the PTT file"
Remote command:
●
BUS<m>:RFFE:SEQuence<n>:SYMBol?
13.17.5
RFFE decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1.
In the "Protocol" dialog > "Setup" tab, enable "Decode".
2.
In the "Protocol" dialog > "Display" tab, select additional result display settings:
"Show decode table" and "Show binary signals".
For a description of the display settings, see also
The instrument captures and decodes the signal according to the standard definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
RFFE (option R&S
RTO6-K550)