Remote control commands
R&S
®
RTO6
2285
User Manual 1801.6687.02 ─ 05
TRIGger<m>:MPHY:DATA:ICONdition
<DataIdxOperator>
Sets the condition for the index. You can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<DataIdxOperator>
EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal | LTHan | LETHan | GTHan | GETHan
Equal, Less than, Less or equal than, Greater Than, Greater or
equal than. These conditions require one pattern to be set with
the corresponding
command.
INRange = RANGe
In range: Set the minimum and maximum value using
*RST:
INRange
TRIGger<m>:MPHY:DATA:IMIN
<DataIndexMin>
Specifies the index minimum, or sets the start value of a range.
Suffix:
<m>
.
1..3
Parameters:
<DataIndexMin>
Range:
1 to 0
Increment: 1
*RST:
1
TRIGger<m>:MPHY:DATA:IMAX
<DataIndexMax>
Sets the end value of the index range if
TRIGger<m>:MPHY:DATA:ICONdition
is
set to
INRange
or
RANge
.
Suffix:
<m>
.
1..3
Parameters:
<DataIndexMax>
Range:
1 to 0
Increment: 1
*RST:
0
TRIGger<m>:MPHY:FSNumber:CONDition
<FSNOperator>
Sets the condition for FSN. You can define an exact value or a value range.
Protocols